Olympic Cool Cap Service Manual



OLYMPIC schematic diagram service manual circuit diagram wiring schema repair instruction guide user manual free pdf download. Service Handbuch Schaltungen Reparaturanleitung Bedienungsanleitungen kostenlos pdf download. This manual contains important information about the assembly, operation and maintenance of this patio heater. General safety information is presented in these first few pages and is also located throughout the manual. Keep this manual for future reference and to educate new users of this product. This manual should be read in.

A medical head-cooling system for newborns is being recalled because of a software problem, which may cause it to fail while treating infants.

Olympic Cool Cap Service Manual Download

The Olympic Cool-Cap System recall was announced by the FDA on December 12, impacting selective head cooling systems for full-term infants suffering from hypoxic-ischemic encephalopathy (HIE), which are manufactured by Natus Medical.

Natus originally sent out an Urgent: Medical Device letter to its customers in May, alerting them that the screen on the Control Module of the device can freeze. This means that despite information being shown on-screen that the caps are working, the infant may not be receiving cooling treatment.

This week the FDA determined that the letter constituted a Class I medical device recall, the highest designation a recall can receive. The Class I designation means that the FDA believes the cooling cap problem creates the reasonable probability that a patient will suffer severe injury or death. However, there have been no injuries or deaths reported in relation to the Olympic Cool-Cap problems.

Olympic Cool Cap Service ManualOlympic Cool Cap Service Manual

The recall affects the Olympic Cool-Cap System, catalog/part number 60010, distributed between January 18, 2007 through April 20, 2012. The recall affects all serial numbers.

The caps are cooling water caps placed on the head of the infant while the body is warmed with radiant warmers. When working properly the caps provide 72 hours of monitored cooling for moderate to severe cases of HIE to prevent or reduce neurologic injury.

Olympic Cool Cap Service Manual Pdf

Customers with questions should contact Natus Technical Service at (888)496-2887.

Olympic Cool Cap Service Manuals